Skin Surface Temperature Probe disposable, sterilized
Product description:
The item is one skin surface temperature probe disposable, sterilized. It's composed of NTC element, polyimide tube, epoxy, patch, wire and connector. They're used for human body surface temperature measurement. Our full spectrum NTC sensors offer many advantages in temperature sensing including miniature size, excellent long-term stability, high accuray and precision.
Product details as below:
JP skin surface temperature probe disposable offer a cost-effective, disposable patient temperature monitoring solution to add to your hospital's infection control program. The sensor helps to reduce the risk of cross contamination between patients.
JP skin surface temperature probes disposable provide accurate measurement of body surface (skin) and core (esophageal/rectal) temperature in adults, infants and neonates.
JP disposable surface temperature probes are compatible with the majority of manufacturers of incubators, warmers and patient monitors.
YSI400 compatible disposable skin surface temperature probes come with Molex Style of 3.5mm Phone Jack connectors. Disposable temperature probes for infant warmers and incubators have different connectors compatible with original manufacturers. The distal end of the temperature probe is terminated with a plastic insert molded, moisture resistant connector.
Reliability test
High temperature Exposure (Storage) |
- Resistance (R25) change should be less than ±5%.
- B-constant (B25/50) change should be less than ±2%.
- No visible damage. |
Leave continuously according to the following table for 1000 hours.
Operating Temp. Range: -40 to +150ºC Type |
150±3ºC |
Operating Temp. Range: -40 to +125ºC Type |
125±3ºC |
Measurement at 24±2 hours after test condition. |
Temperature Cycling |
- Resistance (R25) change should be less than ±5%.
- B-constant(B25/50) change should be less than ±2%.
- No visible damage |
Perform 1000 cycles according to the four heat treatments listed in the following table.
Step |
1 |
2 |
3 |
4 |
Temp.(deg.C) |
-55+0/-3 |
Room Temp. |
125+3/-0 |
Room Temp. |
Time (min.) |
15±3 |
1 |
15±3 |
1 |
Measurement at 24±2 hours after test condition. |
Moisture Resistance |
- Resistance(R25) change should be less than ±5%.
- B-constant(B25/50) change should be less than ±2%.
- No visible damage. |
Apply the 24-hour heat (25 to 65ºC) and humidity (80 to 98%) treatment shown below, 10 consecutive times. Measurement at 24±2 hours after test condition. |
Biased Humidity |
- Resistance (R25) change should be less than ±10%
- B-constant(B25/50) change should be less than ±2%
- No visible damage. |
85±2ºC, 85%RH in air for 1000 hours with Permissive Operating Current. Measurement at 24±2 hours after test conditions |
Operational Life |
- Resistance(R25) change should be less than ±5%
- B-constant (B25/50) change should be less than ±2%
- No visible damage |
85±3ºC in air for 1000 hours with Permissive Operating Current. Measurement at 24±2 hours after test conditions |
External Visual |
No defects of abnormalities |
Visual Inspection. |
Physical Dimension |
Within the specified dimensions |
Using calipers |
![Skin Surface Temperature Probe Disposable, Sterilized](//www.micstatic.com/athena/img/transparent.png)
Adhesive foam evaluated by bio-compatibility can be fixed on measuring position, it is with no irritation to the skin and with better comfort. The foam reflector can effectively insulate ambient temperature and radiant light to ensure normal operation.
FAQ:
Q: Can you show the match behind conversion from % tolerance to actual temperature tolerance?
A: To determine the temperature accuracy simply divide the total deviation (resistance tolerance) by the Alpha for the temperature of interest.
Q: Does the thermistor accuracy specifications include any long term resistance change over time (resistance stability)?
A: No, the accuracy is specified as the accuracy of the sensor shipped. We have no control of the application or the environmental conditions the sensor will be subjected to in the field.
Q: What does % mean when talking about temperature accuracy?
A: Sensor's accuracy can be specified as resistance tolerance or to a temperature accuracy at either a single point at ... over a span.