The item is one skin surface temperature probe single use probe. It's composed of NTC element, polyimide tube, epoxy, patch, wire and connector. They're used for human body surface temperature measurement. Long adapter cable for use with single use temperature probes, reusable converts 2-pin temperature probe connector to 2-pin instrument connector.
JP series skin surface temperature probe single use probe offer a cost-effective, disposable patient temperature monitoring solution to add to your hospital's infection control program. The sensor helps to reduce the risk of cross contamination between patients.
High temperature Exposure (Storage) | - Resistance (R25) change should be less than ±5%. - B-constant (B25/50) change should be less than ±2%. - No visible damage. |
Leave continuously according to the following table for 1000 hours.
|
|||||||||||||||
Temperature Cycling | - Resistance (R25) change should be less than ±5%. - B-constant(B25/50) change should be less than ±2%. - No visible damage |
Perform 1000 cycles according to the four heat treatments listed in the following table.
|
|||||||||||||||
Moisture Resistance | - Resistance(R25) change should be less than ±5%. - B-constant(B25/50) change should be less than ±2%. - No visible damage. |
Apply the 24-hour heat (25 to 65ºC) and humidity (80 to 98%) treatment shown below, 10 consecutive times. Measurement at 24±2 hours after test condition. | |||||||||||||||
Biased Humidity | - Resistance (R25) change should be less than ±10% - B-constant(B25/50) change should be less than ±2% - No visible damage. |
85±2ºC, 85%RH in air for 1000 hours with Permissive Operating Current. Measurement at 24±2 hours after test conditions | |||||||||||||||
Operational Life | - Resistance(R25) change should be less than ±5% - B-constant (B25/50) change should be less than ±2% - No visible damage |
85±3ºC in air for 1000 hours with Permissive Operating Current. Measurement at 24±2 hours after test conditions | |||||||||||||||
External Visual | No defects of abnormalities | Visual Inspection. | |||||||||||||||
Physical Dimension | Within the specified dimensions | Using calipers |